General operation of solid phase extraction
Reverse extraction column operation
Normal phase extraction column operation
Ion exchange extraction column
Extracted from polar matrix
Non-polar samples
Extracted from non-polar matrix
Polar samples
Extract charged samples
activation
Rinse the extraction column with 3-5 ml of methanol;
Rinse the extraction column with 3-5 ml of deionized water (or buffer) and keep the extraction column moist
3-5ml non-polar solvent
Elution extraction column
Rinse the extraction column with 3-5 ml of deionized water and / or low ionic strength buffer
Loading
Sample at the appropriate flow rate
Evenly through the extraction column
With a low flow rate, the sample is evenly passed through the extraction column
Cleaning
An extraction column that elutes the adsorbed sample with 5 ml of a solvent of appropriate polarity (commonly used are pure water, buffer, water / organic solvent mixture)
Elute the extraction column with 5 ml of a suitable non-polar solvent
Rinse the extraction column with 5 ml of deionized water and / or low ionic strength buffer
Elution
The test sample is eluted in batches with 1-5 ml non-polar solvent to the collection container
The test sample is eluted in batches with 1-5 ml of polar solvent to the collection container
The test sample is eluted with 1-5 ml of high ionic strength buffer to the collection container